LK-J63 Gutta Cutter

  • LK-J63 Gutta Cutter
  • LK-J63 Gutta Cutter
  • LK-J63 Gutta Cutter
  • LK-J63 Gutta Cutter
  • LK-J63 Gutta Cutter
  • LK-J63 Gutta Cutter
LK-J63 Gutta CutterLK-J63 Gutta CutterLK-J63 Gutta CutterLK-J63 Gutta CutterLK-J63 Gutta CutterLK-J63 Gutta Cutter

LK-J63 Gutta Cutter

LK-J63 Gutta Cutter

1. Structural and composition: 
Made of high quality stainless steel and aluminum alloy Material


2. Scope of application: 
It is suitable for cutting off the Gutta-Percha￾Points outside the mouth for root canal filling.


3. Operation instruction:
3.1 This device is only suitable for its corresponding scope of application. It should not 
be used for other purposes. Using it for other purposes will damage the product. It should be 
operated and used by the professionals. New products must be cleaned, disinfected and 
sterilized before use. When taking a reading, the eye's line of sight should be perpendicular to the 
scale surface.
3.2 This instrument is mainly divided into two parts, a measuring ruler with a scale and a gutta-percha point hole, and a sliding block for trimming the gutta-percha point.
3.3 Demonstration of use: Put the gutta-percha point into the # 25 hole,

make sure that the gutta-percha point is in place, push the sliding block, 

remove the excess part of the gutta-percha point, and you will get a #25 gutta-percha tip.


4. Maintenance:
4.1 Before use, remove the residue with ultrasonic and sterilize it;
4.2 Before putting the Gauge for Gutta-Percha￾Points into the sterilization bag, it should be completely dry;
4.3 After sterilizing the Gauge for Gutta-Percha￾Points, do not contact it with other equipment to avoid contamination;
4.4 Do not perform chemical sterilization;
4.5 The maximum reusable times are 500 times.


5. Sterilization method: 
Select the temperature, pressure and time of the autoclave sterilizer: 134℃ ,

2.0bar-2.3bar (0.20MPa-0.23MPa), 4min.


6. Cautions:
6.1 Please clean, disinfect and sterilize this product before using it for the first time.
6.2 This product should be cleaned and sterilized in time after use.
6.3 Avoid placing this product with chemicals.
6.4 Do not use this product if it is found to be rusted, damaged or the scale is blurred.
6.5 This product should be stored in a well-ventilated, moisture-proof room without corrosive gas.


7. Storage condition: 
This product should be stored in a clean, dry and ventilated indoor place where is with relative 
humidity 10% ~ 93%, with atmospheric pressure 70kPa ~ 106kPa, temperature -20 °C ~ +55 ° C 
and non-corrosive gas.